User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Trig-Tek 203M:Charge Amplifier DATE: 05-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 259 CONFIGURATION: Fluke 5700A CONFIGURATION: Datron 1281 STANDARD: DTB 1000pF ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD DTB 1000pF 1.004 HEAD * INITIAL CONDITIONS * 1.005 DISP This test requires the following Standard equipment: 1.005 DISP DTB 1000pF Standard Capacitor - DTB# 19-18 1.006 IEEE [@1281]*RST 1.007 IEEE [@1281]ACV AUTO 1.008 HEAD {* Acceleration Output *} 1.009 JMP 2.001 1.010 EVAL 2.001 DISP Set the UUT as follows: 2.001 DISP CAL/OPER .............................. OPER 2.001 DISP SENSITIVITY switch .................... 1-11 2.001 DISP SENSITIVITY dial ...................... 10.0 2.001 DISP PEAK/RMS .............................. RMS 2.001 DISP SE/DIFF ............................... SE 2.002 DISP Set the UUT as follows: 2.002 DISP UNITS ................................. g'S 2.002 DISP FULL SCALE ............................ 100 2.002 DISP INPUT ................................. pC/g 2.002 DISP OUTPUT ................................ 10 mV/g 2.002 DISP LP FILTER ............................. MED 2.003 DISP Connect the 5700A OUTPUT to the ACCEL INPUT (top) 2.003 DISP connector on the Test card through the 1000pF standard 2.003 DISP capacitor. 2.004 DISP Connect the Datron 1281 HI/LO input to the UUT 2.004 DISP ACCEL OUT BNC connector. 2.005 5700 0.70700V 139.9H S 2W 2.006 ACC .70700V 0.001U 2.007 IEEE [@1281]RDG?[D5000][I] 2.008 MEME 2.009 MEMC V 3.000% 10mV/g 3.001 DISP Set the UUT as follows: 3.001 DISP OUTPUT ................................ 100 mV/g 3.002 5700 0.707V 139.9H S 2W 3.003 ACC 7.0700V 0.001U 3.004 IEEE [@1281]RDG?[D5000][I] 3.005 MEME 3.006 MEMC V 3.000% 100mV/g 4.001 HEAD {} 4.002 HEAD {* Meter Tests *} 4.003 JMP 5.001 4.004 EVAL 5.001 DISP Set the UUT as follows: 5.001 DISP OUTPUT ................................ 10 mV/g 5.002 5700 0.70700V 139.9H S 2W 5.003 MEMI Enter the UUT meter reading. 5.004 ACC 70.7g's 0.1U 5.005 MEME 5.006 MEMC g's 2.0U rms 6.001 5700 0.70700V 139.9H S 2W 6.002 DISP Set the UUT as follows: 6.002 DISP PEAK/RMS .............................. PEAK 6.003 MEMI Enter the UUT meter reading. 6.004 ACC 100.0g's 0.1U 6.005 MEME 6.006 MEMC g's 2.0U peak 7.001 HEAD {} 7.002 HEAD {* Range Tests *} 7.003 JMP 8.001 7.004 EVAL 8.001 DISP Set the UUT controls as follows: 8.001 DISP FULL SCALE ............................ 1000 8.002 5700 7.0700V 139.9H S 2W 8.003 ACC 1000g's 1U 8.004 MEMI Enter the UUT meter reading. 8.005 MEME 8.006 MEMC g's 20U 9.001 5700 7.0700V 139.9H S 2W 9.002 ACC 7.0700V 0.01% 9.003 IEEE [@1281]RDG?[D5000][I] 9.004 MEME 9.005 MEMC V 3.0% 10.001 HEAD * Range Tests * 10.002 JMP 11.001 10.003 EVAL 11.001 DISP Set the UUT controls as follows: 11.001 DISP FULL SCALE ............................ 100 11.002 5700 0.707000V 139.9H S 2W 11.003 ACC 100.0g's 0.01U 11.004 MEMI Enter the UUT meter reading. 11.005 MEME 11.006 MEMC g's 2.0U 12.001 5700 0.707000V 139.9H S 2W 12.002 ACC .70700V 0.01% 12.003 IEEE [@1281]RDG?[D5000][I] 12.004 MEME 12.005 MEMC V 3% 13.001 HEAD * Range Tests * 13.002 JMP 14.001 13.003 EVAL 14.001 DISP Set the UUT controls as follows: 14.001 DISP FULL SCALE ............................ 10 14.002 5700 70.700mV 139.9H S 2W 14.003 ACC 10.00g's 0.01U 14.004 MEMI Enter the UUT meter reading. 14.005 MEME 14.006 MEMC g's 0.2U 15.001 5700 70.700mV 139.9H S 2W 15.002 ACC 70.700mV 0.01% 15.003 IEEE [@1281]RDG?[D5000][I] 15.004 MEM* 1000 15.005 MEME 15.006 MEMC mV 3% 16.001 IEEE [@1281]DCV AUTO 16.002 HEAD {* DC Output *} 16.003 JMP 17.001 16.004 EVAL 17.001 5700 70.700mV 139.9H S 2W 17.002 DISP Connect the Datron 1281 to the UUT DC OUTPUT BNC. 17.003 ACC 10.000V 0.001% 17.004 IEEE [@1281]RDG?[D1000][I] 17.005 MEME 17.006 MEMC V 0.2U 18.001 HEAD {* Displacement Output Tests *} 18.002 JMP 19.001 18.003 EVAL 19.001 DISP Set the UUT controls as follows: 19.001 DISP UNITS ................................. MILS 19.001 DISP FULL SCALE ............................ 100 19.002 IEEE [@1281]ACV AUTO 19.003 5700 0.70700V 139.9H S 2W 19.004 DISP Connect the Datron 1281 to the UUT DISP OUT BNC. 19.005 ACC 100mils 0.1U 19.006 MEMI Enter the UUT meter reading. 19.007 MEME 19.008 MEMC mils 3U 20.001 5700 0.70700V 139.9H S 2W 20.002 ACC 1.0000V 0.001% 20.003 IEEE [@1281]RDG?[D5000][I] 20.004 MEME 20.005 MEMC V 0.03U 21.001 HEAD {* SENSITIVITY Dial Test *} 21.002 JMP 23.003 21.003 EVAL 22.001 JMP 23.001 22.002 EVAL 23.001 DISP Reset the UUT SENSITIVITY dial to 10.0 23.002 JMP 23.005 23.003 DISP Set the UUT controls as follows: 23.003 DISP UNITS ................................. g's 23.004 DISP Connect the Datron 1281 to the UUT ACCEL OUT BNC. 23.005 5700 0.70700V 139.9H S 2W 23.006 IEEE [@1281]RDG?[D5000][I] 23.007 MEME >1 23.008 5700 70.700mV 139.9H S 2W 23.009 DISP Set the UUT controls as follows: 23.009 DISP SENSITIVITY dial ...................... 1.0 23.010 ACC .70700v 0.001% 23.011 IEEE [@1281]RDG?[D5000][I] 23.012 MEME 23.013 MEME <1 23.014 MEMC V 3% 24.001 HEAD {* SENSITIVITY Range Switch Test *} 24.002 JMP 25.001 24.003 EVAL 25.001 DISP Set the UUT controls as follows: 25.001 DISP SENSITIVITY dial ...................... 10.0 25.002 5700 0.70700V 139.9H S 2W 25.003 IEEE [@1281]RDG?[D5000][I] 25.004 MEME >1 25.005 DISP Set the UUT controls as follows: 25.005 DISP SENSITIVITY switch .................... 10-110 25.006 5700 7.0700V 139.9H S 2W 25.007 ACC .70700V 0.01% 25.008 IEEE [@1281]RDG?[D5000][I] 25.009 MEME 25.010 MEME <1 25.011 MEMC V 3% 26.001 HEAD {* LP FILTER Tests *} 26.002 HEAD {HIGH} 26.003 HEAD * LP FILTER Tests * 26.004 JMP 27.001 26.005 EVAL 27.001 DISP Set the UUT controls as follows: 27.001 DISP LP FILTER ............................. HIGH 27.001 DISP SENSITIVITY switch .................... 1-11 27.002 MESS Adjust the Stimulus for a reading of 100.0 on 27.002 MESS the UUT meter. 27.003 5700 0.707V 139.9H N 2W 27.004 MESS Adjust the Stimulus for a reading of 70.7 on the 27.004 MESS UUT meter. 27.005 5700 30.00kH 20.0% V 2W #27.004 5700 27.00kH V S 2W #27.005 MEMI Enter the UUT meter reading. #27.006 MATH mem1 = 70.7 #27.007 MEME #27.008 MEMC g's 1.0U HIGH 28.001 HEAD {MED} 28.002 HEAD * LP FILTER Tests * 28.003 JMP 29.001 28.004 EVAL 29.001 DISP Set the UUT controls as follows: 29.001 DISP LP FILTER ............................. MED 29.002 MESS Adjust the Stimulus for a reading of 100.0 on 29.002 MESS the UUT meter. 29.003 5700 0.707V 139.9H N 2W 29.004 MESS Adjust the Stimulus for a reading of 70.7 on the 29.004 MESS UUT meter. 29.005 5700 10.000kH 20.0% V 2W #29.004 5700 9.70kH V S 2W #29.005 MEMI Enter the UUT meter reading. #29.006 MATH mem1 = 70.7 #29.007 MEME #29.008 MEMC g's 1.0U MED 30.001 HEAD {LOW} 30.002 HEAD * LP FILTER Tests * 30.003 JMP 31.001 30.004 EVAL 31.001 DISP Set the UUT controls as follows: 31.001 DISP LP FILTER ............................. LOW 31.002 MESS Adjust the Stimulus for a reading of 100.0 on 31.002 MESS the UUT meter. 31.003 5700 0.707V 139.9H N 2W 31.004 MESS Adjust the Stimulus for a reading of 70.7 on the 31.004 MESS UUT meter. 31.005 5700 3.000kH 20.0% V 2W #31.004 5700 2.70kH V S 2W #31.005 MEMI Enter the UUT meter reading. #31.006 MATH mem1 = 70.7 #31.007 MEME #31.008 MEMC g's 1.0U LOW 32.001 HEAD {* CAL Tests *} 32.002 JMP 33.001 32.003 EVAL 33.001 DISP Set the UUT controls as follows: 33.001 DISP OPER/CAL .............................. CAL 33.002 DISP Connect the Datron 1281 to the UUT front panel 33.002 DISP INPUT/MONITOR BNC. 33.003 HEAD [27][91]7m[27][91]5mPlease Wait ...... Long Gate Time 33.004 IEEE [@1281]GATE FAST_OFF[D1000] 33.005 IEEE [@1281]RDG?[D5000][I] 33.006 HEAD 33.007 ACC 1.0000V 0.01% 33.008 MEME 33.009 MEMC V 0.050U 34.001 HEAD [27][91]7m[27][91]5mPlease Wait ...... Long Gate Time 34.002 IEEE [@1281]FREQ?[D10000][I] 34.003 ACC 139.90H 0.01% 34.004 HEAD 34.005 MEME 34.006 MEMC H 2.0U 35.001 IEEE [@1281]*RST 35.002 END